Recall:arbitrary[:Start] - Agilent Technologies InfiniiVision 3000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 3000 X-Series:
Table of Contents

Advertisement

:RECall:ARBitrary[:STARt]

Command Syntax
:RECall:ARBitrary:[STARt] [<file_spec>][, <column>]
<file_spec> ::= {<internal_loc> | <file_name>}
<column> ::= Column in CSV file to load. Column number starts from 1.
<internal_loc> ::= 0-3; an integer in NR1 format
<file_name> ::= quoted ASCII string
The :RECall:ARBitrary:[STARt] command recalls an arbitrary waveform.
If a file extension is provided as part of a specified <file_name>, it must be ".csv".
N O T E
For internal locations, the <column> parameter is ignored.
For external (USB storage device) files, the column parameter is optional.
If no <column> parameter is entered, and it is a 2- column file, the 2nd
column (assumed to be voltage) is automatically be selected. If the
<column> parameter is entered, and that column does not exist in the file,
the operation fails.
When recalling arbitrary waveforms (from an external USB storage device)
that were not saved from the oscilloscope, be aware that the oscilloscope
uses a maximum of 8192 points for an arbitrary waveform. For more
efficient recalls, make sure your arbitrary waveforms are 8192 points or
less.
See Also
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
(see
page
1138)
"Introduction to :RECall Commands"
":RECall:FILename"
on page 600
":SAVE:ARBitrary[:STARt]"
on page 598
on page 608
25
:RECall Commands
599

Advertisement

Table of Contents
loading

Table of Contents