Root (:) Commands
:AER (Arm Event Register)
Query Syntax
:AER?
The AER query reads the Arm Event Register. After the Arm Event
Register is read, it is cleared. A "1" indicates the trigger system is in the
armed state, ready to accept a trigger.
The Armed Event Register is summarized in the Wait Trig bit of the
Operation Status Event Register. A Service Request can be generated when
the Wait Trig bit transitions and the appropriate enable bits have been set
in the Operation Status Enable Register (OPEE) and the Service Request
Enable Register (SRE).
Return Format
<value><NL>
<value> ::= {0 | 1}; an integer in NR1 format.
See Also
192
(see
page
1138)
"Introduction to Root (:) Commands"
":OPEE (Operation Status Enable Register)"
":OPERegister:CONDition (Operation Status Condition Register)"
page 207
":OPERegister[:EVENt] (Operation Status Event Register)"
"*STB (Read Status Byte)"
"*SRE (Service Request Enable)"
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide