System:preset - Agilent Technologies InfiniiVision 3000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 3000 X-Series:
Table of Contents

Advertisement

29
:SYSTem Commands

:SYSTem:PRESet

Command Syntax
:SYSTem:PRESet
The :SYSTem:PRESet command places the instrument in a known state.
This is the same as pressing the [Default Setup] key or [Save/Recall] >
Default/Erase > Default Setup on the front panel.
When you perform a default setup, some user settings (like preferences)
remain unchanged. To reset all user settings to their factory defaults, use
the *RST command.
Reset conditions are:
Acquire Menu
Mode
Averaging
# Averages
Analog Channel Menu
Channel 1
Channel 2
Volts/division
Offset
Coupling
Probe attenuation
Vernier
Invert
BW limit
Impedance
Units
Skew
Cursor Menu
Source
858
(see
page
1138)
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
Normal
Off
8
On
Off
5.00 V
0.00
DC
10:1
Off
Off
Off
1 M Ohm (cannot be changed)
Volts
0
Channel 1