:POWer:HARMonics:APPLy

Command Syntax
:POWer:HARMonics:APPLy
The :POWer:HARMonics:APPLy command applies the current harmonics
analysis.
Switching power supplies draw a range of harmonics from the AC mains.
Standard limits are set for these harmonics because these harmonics can
travel back to the supply grid and cause problems with other devices on
the grid.
Use the Current Harmonics analysis to test a switching power supply's
current harmonics to pre- compliance standard of IEC61000- 3- 2 (Class A,
B, C, or D). The analysis presents up to 40 harmonics.
See Also
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
(see
page
1138)
":POWer:HARMonics:DATA"
":POWer:HARMonics:DISPlay"
":POWer:HARMonics:FAILcount"
":POWer:HARMonics:LINE"
":POWer:HARMonics:POWerfactor"
":POWer:HARMonics:STANdard"
":POWer:HARMonics:STATus"
":POWer:HARMonics:RUNCount"
":POWer:HARMonics:THD"
on page 538
on page 539
on page 540
on page 541
on page 542
on page 544
on page 545
on page 543
on page 546
24
:POWer Commands