Measure:efficiency - Agilent Technologies InfiniiVision 3000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 3000 X-Series:
Table of Contents

Advertisement

21
:MEASure Power Commands

:MEASure:EFFiciency

Command Syntax
:MEASure:EFFiciency
The :MEASure:EFFiciency command installs an efficiency (output power /
input power) measurement on screen.
Before sending this command or query, you must specify the channels
probing the input voltage, input current, output voltage, and output
current (using the :POWer:SIGNals:SOURce:VOLTage<i> and
:POWer:SIGNals:SOURce:CURRent<i> commands) and you must perform the
automated signals setup (using the :POWer:SIGNals:AUTosetup EFFiciency
command).
Query Syntax
:MEASure:EFFiciency?
The :MEASure:EFFiciency query returns the measured efficiency as a
percent value.
Return Format
<return_value><NL>
<return_value> ::= percent value in NR3 format
See Also
476
(see
page
1138)
":POWer:SIGNals:SOURce:VOLTage<i>"
":POWer:SIGNals:SOURce:CURRent<i>"
":POWer:SIGNals:AUTosetup"
":POWer:EFFiciency:APPLy"
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
on page 582
on page 581
on page 564
on page 535