16
:FUNCtion Commands
Table 88 :FUNCtion Commands Summary (continued)
Command
:FUNCtion:SOURce2
<source> (see
page
354)
:FUNCtion:TRENd:MEASu
rement <type> (see
page
355)
Introduction to
The FUNCtion subsystem controls the math functions in the oscilloscope.
:FUNCtion
As selected by the OPERation command, these math functions are
Commands
available:
• Operators:
• Transforms:
With the DSOX3ADVMATH advanced math measurements license, these
additional math functions are available:
• Operators:
• Transforms:
326
Query
:FUNCtion:SOURce2?
(see
page
354)
:FUNCtion:TRENd:MEASu
rement? (see
page
355)
ADD
SUBTract
MULTiply
Operators perform their function on two analog channel sources.
DIFF — Differentiate
INTegrate — The INTegrate:IOFFset command lets you specify a DC
offset correction factor.
FFT — The SPAN, CENTer, VTYPe, and WINDow commands are used
for FFT functions. When FFT is selected, the horizontal cursors
change from time to frequency (Hz), and the vertical cursors change
from volts to decibel (dB).
SQRT — Square root
Transforms operate on a single analog channel source or on a g(t)
function that is the addition, subtraction, or multiplication of analog
channel sources (specified by the GOFT commands).
DIVide
LINear — Ax + B — The LINear commands set the gain (A) and offset
(B) values for this function.
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
Options and Query Returns
<source> ::= {CHANnel<n> | NONE}
<n> ::= {{1 | 2} | {3 | 4}} for
4ch models, depending on SOURce1
selection
<n> ::= {1 | 2} for 2ch models
<type> ::= {VAVerage | ACRMs |
VRATio | PERiod | FREQuency |
PWIDth | NWIDth | DUTYcycle |
RISetime | FALLtime}