:TIMebase:WINDow:POSition

Command Syntax
:TIMebase:WINDow:POSition <pos value>
<pos value> ::= time from the trigger event to the zoomed (delayed)
The :TIMebase:WINDow:POSition command sets the horizontal position in
the zoomed (delayed) view of the main sweep. The main sweep range and
the main sweep horizontal position determine the range for this command.
The value for this command must keep the zoomed view window within
the main sweep range.
Query Syntax
:TIMebase:WINDow:POSition?
The :TIMebase:WINDow:POSition? query returns the current horizontal
window position setting in the zoomed view.
Return Format
<value><NL>
<value> ::= position value in seconds
See Also
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
(see
page
1138)
view reference point in NR3 format