Timebase:window:scale - Agilent Technologies InfiniiVision 3000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 3000 X-Series:
Table of Contents

Advertisement

:TIMebase:WINDow:SCALe

Command Syntax
:TIMebase:WINDow:SCALe <scale_value>
<scale_value> ::= scale value in seconds in NR3 format
The :TIMebase:WINDow:SCALe command sets the zoomed (delayed)
window horizontal scale (seconds/division). The main sweep scale
determines the range for this command. The maximum value is one half of
the :TIMebase:SCALe value.
Query Syntax
:TIMebase:WINDow:SCALe?
The :TIMebase:WINDow:SCALe? query returns the current zoomed window
scale setting.
Return Format
<scale_value><NL>
<scale_value> ::= current seconds per division for the zoomed window
See Also
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
(see
page
1138)
"Introduction to :TIMebase Commands"
":TIMebase:RANGe"
on page 869
":TIMebase:POSition"
":TIMebase:SCALe"
on page 871
":TIMebase:WINDow:RANGe"
on page 866
on page 868
on page 874