:MEASure Commands

:MEASure:DUAL:VAVerage

Overview
This measurement is available with the N2820A high sensitivity current
probe when both the Primary and Secondary probe cables are used. This
measurement joins the Zoom In waveform data below the probe's clamp
level with Zoom Out waveform data above the probe's clamp level to create
the waveform on which the measurement is made.
Command Syntax
:MEASure:DUAL:VAVerage [<interval>][,<source1>][,<source2>]
<interval> ::= {CYCLe | DISPlay}
<source1>,<source2> ::= CHANnel<n> with N2820A probe connected
<n> ::= 1 to (# analog channels) in NR1 format
The :MEASure:DUAL:VAVerage command installs a screen measurement
and starts an average value measurement.
The <interval> option lets you specify the measurement interval: either an
integral number of cycles, or the full screen. If <interval> is not specified,
DISPlay is implied.
If the optional source parameter(s) are specified, the currently specified
source(s) are modified.
Query Syntax
:MEASure:DUAL:VAVerage? [<interval>][,<source1>][,<source2>]
The :MEASure:DUAL:VAVerage? query returns the average value
measurement.
Return Format
<value><NL>
<value> ::= calculated average value in NR3 format
See Also
420
(see
page
1138)
":MEASure:DUAL:CHARge"
":MEASure:DUAL:VAMPlitude"
":MEASure:DUAL:VBASe"
":MEASure:DUAL:VPP"
":MEASure:DUAL:VRMS"
"Introduction to :MEASure Commands"
":MEASure:SOURce"
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
on page 418
on page 419
on page 421
on page 422
on page 423
on page 406
on page 444