:MARKer:MODE

Command Syntax
:MARKer:MODE <mode>
<mode> ::= {OFF | MEASurement | MANual | WAVeform}
The :MARKer:MODE command sets the cursors mode:
• OFF — removes the cursor information from the display.
• MANual — enables manual placement of the X and Y cursors.
• MEASurement — cursors track the most recent measurement.
• WAVeform — the Y1 cursor tracks the voltage value at the X1 cursor of
Query Syntax
:MARKer:MODE?
The :MARKer:MODE? query returns the current cursors mode.
Return Format
<mode><NL>
<mode> ::= {OFF | MEAS | MAN | WAV}
See Also
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
(see
page
1138)
If the front- panel cursors are off, or are set to the front- panel Hex or
Binary mode, setting :MARKer:MODE MANual will put the cursors in
the front- panel Normal mode.
Setting the mode to MEASurement sets the marker sources
(:MARKer:X1Y1source and :MARKer:X2Y2source) to the measurement
source (:MEASure:SOURce). Setting the measurement source remotely
always sets the marker sources.
the waveform specified by the X1Y1source, and the Y2 cursor does the
same for the X2 cursor and its X2Y2source.
"Introduction to :MARKer Commands"
":MARKer:X1Y1source"
":MARKer:X2Y2source"
":MEASure:SOURce"
":MARKer:X1Position"
":MARKer:X2Position"
":MARKer:Y1Position"
":MARKer:Y2Position"
on page 380
on page 383
on page 385
on page 444
on page 382
on page 384
on page 389
on page 390
19
:MARKer Commands
381