27
:SBUS<n> Commands
:SBUS<n>:SPI:TRIGger:TYPE
Command Syntax
:SBUS<n>:SPI:TRIGger:TYPE <value>
<value> ::= {MOSI | MISO}
The :SBUS<n>:SPI:TRIGger:TYPE command specifies whether the SPI
trigger will be on the MOSI data or the MISO data.
When triggering on MOSI data, the data value is specified by the
:SBUS<n>:SPI:TRIGger:PATTern:MOSI:DATA and
:SBUS<n>:SPI:TRIGger:PATTern:MOSI:WIDTh commands.
When triggering on MISO data, the data value is specified by the
:SBUS<n>:SPI:TRIGger:PATTern:MISO:DATA and
:SBUS<n>:SPI:TRIGger:PATTern:MISO:WIDTh commands.
Query Syntax
:SBUS<n>:SPI:TRIGger:TYPE?
The :SBUS<n>:SPI:TRIGger:TYPE? query returns the current SPI trigger
type setting.
Return Format
<value><NL>
<value> ::= {MOSI | MISO}
See Also
•
•
•
•
•
•
•
•
752
(see
page
1138)
"Introduction to :TRIGger Commands"
":SBUS<n>:SPI:SOURce:MOSI"
":SBUS<n>:SPI:SOURce:MISO"
":SBUS<n>:SPI:TRIGger:PATTern:MISO:DATA"
":SBUS<n>:SPI:TRIGger:PATTern:MOSI:DATA"
":SBUS<n>:SPI:TRIGger:PATTern:MISO:WIDTh"
":SBUS<n>:SPI:TRIGger:PATTern:MOSI:WIDTh"
":TRIGger:MODE"
on page 886
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
on page 877
on page 747
on page 746
on page 748
on page 750
on page 749
on page 751