:SBUS<n>:A429:FORMat
Command Syntax
:SBUS<n>:A429:FORMat <format>
<format> ::= {LDSDi | LDSSm | LDATa}
The :SBUS<n>:A429:FORMat command specifies the word decode format:
• LDSDi:
• LDSSm:
• LDATa:
Query Syntax
:SBUS<n>:A429:FORMat?
The :SBUS<n>:A429:FORMat? query returns the current ARINC 429 word
decode format setting.
Return Format
<format><NL>
<format> ::= {LDSD | LDSS | LDAT}
Errors
See Also
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
(see
page
1138)
Label - 8 bits.
SDI - 2 bits.
Data - 19 bits.
SSM - 2 bits.
Label - 8 bits.
Data - 21 bits.
SSM - 2 bits.
Label - 8 bits.
Data - 23 bits.
"- 241, Hardware missing"