6
Root (:) Commands
:TER (Trigger Event Register)
Query Syntax
:TER?
The :TER? query reads the Trigger Event Register. After the Trigger Event
Register is read, it is cleared. A one indicates a trigger has occurred. A
zero indicates a trigger has not occurred.
The Trigger Event Register is summarized in the TRG bit of the Status
Byte Register (STB). A Service Request (SRQ) can be generated when the
TRG bit of the Status Byte transitions, and the TRG bit is set in the
Service Request Enable register. The Trigger Event Register must be
cleared each time you want a new service request to be generated.
Return Format
<value><NL>
<value> ::= {1 | 0}; a 16-bit integer in NR1 format.
See Also
224
(see
page
1138)
"Introduction to Root (:) Commands"
"*SRE (Service Request Enable)"
"*STB (Read Status Byte)"
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
on page 190
on page 180
on page 182