:DISPlay Commands
Table 85 :DISPlay Commands Summary (continued)
Command
:DISPlay:PERSistence
<value> (see
page
307)
:DISPlay:VECTors {1 |
ON} (see
page
308)
Introduction to
The DISPlay subsystem is used to control the display storage and retrieval
:DISPlay
of waveform data, labels, and text. This subsystem allows the following
Commands
actions:
• Clear the waveform area on the display.
• Turn vectors on or off.
• Set waveform persistence.
• Specify labels.
• Save and Recall display data.
Reporting the Setup
Use :DISPlay? to query the setup information for the DISPlay subsystem.
Return Format
The following is a sample response from the :DISPlay? query. In this case,
the query was issued following a *RST command.
:DISP:LAB 0;VECT 1;PERS MIN
298
Query
:DISPlay:PERSistence?
(see
page
307)
:DISPlay:VECTors?
(see
page
308)
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
Options and Query Returns
<value> ::= {MINimum | INFinite |
<time>}
<time> ::= seconds in in NR3
format from 100E-3 to 60E0
1