Video Trigger Level Frequency

:TRIGger[:SEQuence]:VIDeo:LEVel:FREQuency <freq>
:TRIGger[:SEQuence]:VIDeo:LEVel:FREQuency?
This command is used to adjust the Video Trigger level when in FM Demod, and
Demod View is on.
Default Unit:
Remarks:
NOTE
Trigger Delay is not available in Video trigger mode, so turning Video on turns off
Trigger Delay, but preserves the value of Trigger Delay.
Chapter 5
Hz
Video is adjusted using this command, but must also be selected
using the command :TRIGger[:SEQuence]:SOURce
VIDeo . When not in FM Demod, the Video Trigger level is
adjusted/queried using the command
:TRIGger[:SEQuence]:VIDeo:LEVel <ampl> .