Advertisement

RF Burst Relative Mode Setting

:TRIGger[:SEQuence]:RFBurst:LEVel:RELative <rel_ampl>
:TRIGger[:SEQuence]:RFBurst:LEVel?
Set the trigger level when using the RF burst (wideband) trigger. The value is
relative to the peak of the burst. RF burst is also known as RF envelope.
This command is available only when Option B7E (RF Communications
Hardware) is installed.
Factory Preset
− 6 dB
and *RST:
− 45 dB to 0 dB
Range:
Default Unit:
dB
History:
Added with firmware revision A.08.00.
Front Panel
Access:
Trig, More 1 of 2, RF Burst Setup
RF Burst Mode Setting
:TRIGger[:SEQuence]:RFBurst:LEVel:TYPE ABSolute|RELative
:TRIGger[:SEQuence]:RFBurst:LEVel:TYPE?
Select the trigger level of the RF burst envelope signal. The level can be set to
either a level relative to the peak of the burst (RELative mode) or to an absolute
level (ABSolute mode).
This command is available only when Option B7E (RF Communications
Hardware) is installed.
Factory Preset
and *RST:
Relative
History:
Added with firmware revision A.08.00.
Front Panel
Access:
Trig, More 1 of 2, RF Burst Setup
RF Burst Narrow Pulse Discriminator
:TRIGger[:SEQuence]:RFBurst:NPDiscrim[:STATe] OFF|ON|0|1
:TRIGger[:SEQuence]:RFBurst:NPDiscrim?
The narrow pulse discriminator prevents undesirable re-triggering on multiple
peaks when the RF envelope has significant ripple.