Rf Burst Frequency Selectivity - Agilent Technologies ESA Series User's/Programmer's Reference

Core spectrum analyzer functions
Hide thumbs Also See for ESA Series:
Table of Contents

Advertisement

Remote Command Reference
TRIGger Subsystem
The effective trigger offset value will be re-calculated whenever
any of these parameters change.
Front Panel
Access:
Trig, Trig Offset

RF Burst Frequency Selectivity

:TRIGger[:SEQuence]:RFBurst:FSELectivity[:STATe] OFF|ON|0|1
:TRIGger[:SEQuence]:RFBurst:FSELectivity?
The RF burst circuit can sample the 21.4 MHz IF either before or after RBW
filtering. When frequency selectivity is On, the RBW-filtered IF signal is used,
increasing the frequency selectivity.
This command is available only when Option B7E (RF Communications
Hardware) is installed.
Factory Preset
and *RST:
Off
History:
Added with firmware revision A.08.00.
Front Panel
Access:
Trig, More 1 of 2, RF Burst Setup, Frequency Selectivity
RF Burst Absolute Mode Setting
:TRIGger[:SEQuence]:RFBurst:LEVel:ABSolute <ampl>
:TRIGger[:SEQuence]:RFBurst:LEVel?
Select the trigger level of the RF burst envelope signal. The value can be set to an
absolute level, with the units displayed in the current y-axis units.
This command is available only when Option B7E (RF Communications
Hardware) is installed.
Factory Preset
0 dBm
and *RST:
− 100 dBm to 100 dBm
Range:
Default Unit:
dBm
History:
Added with firmware revision A.08.00.
Front Panel
Access:
Trig, More 1 of 2, RF Burst Setup
382
Chapter 5

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents