Front Panel
Access:
FREQUENCY, Scale Type

Sweep Time

[:SENSe]:SWEep:TIME <time>
[:SENSe]:SWEep:TIME?
Specifies the time in which the instrument sweeps the display.
Factory Preset
and *RST:
ESA E4401B, E4411B: 4 ms
ESA E4402B, E4403B: 5 ms
ESA E4404B: 16.75 ms
ESA E4405B: 33 ms
ESA E4407B, E4408B: 265 ms
Range:
The range depends upon the installed options, number of sweep
points, and firmware revision of your instrument. See "Sweep
Time Range" in the Specifications Guide for details.
Default Unit:
seconds
Remarks:
A span value of 0 Hz causes the analyzer to enter zero span
mode. In zero span the X-axis represents time rather than
frequency. In this mode, the sweep time may be set to faster
values when Options AYX or B7D are installed.
Front Panel
Access:
Sweep, Sweep Time Auto Man
Automatic Sweep Time
[:SENSe]:SWEep:TIME:AUTO OFF|ON|0|1
[:SENSe]:SWEep:TIME:AUTO?
Automatically selects the fastest sweep time for the current settings.
Factory Preset
and *RST:
On
History:
This command function changed with firmware revision
A.08.00. With :AUTO ON in zero span, an error will be
generated.
Chapter 5
Remote Command Reference
[:SENSe]:SWEep Subsection