Front-Panel Key Reference

Input/Output

Input/Output
Sets the input impedance for voltage-to-power conversions. The
Input Z Corr
impedance you select is for computational purposes only, since the
50Ω 75Ω
actual impedance is set by internal hardware to 50 Ω (except for
Option 1DP). The default is 50 Ω (75 Ω with Option 1DP). Setting the
computational input impedance to 75 Ω is useful when using a 75 Ω to
50 Ω adapter to measure a 75 Ω device on an analyzer having a 50 Ω
input impedance.
Key Access:
Only available in Agilent models E4402B or E4407B with Option UKB,
Coupling
and E4404B or E4405B. Specifies alternating current (ac) or direct
AC DC
current (dc) coupling at the analyzer input. Selecting ac coupling blocks
any dc voltage at the analyzer input, but also decreases the frequency
range of the analyzer. Input coupling is set to ac by an instrument
preset (With
apply only when coupling is set to dc. Refer to the appropriate
amplitude specifications and characteristics for your analyzer.
Table 2-1
Selecting Input Coupling
ESA Model #
E4402B with
Option UKB
E4404B
E4404B with
Option UKB
E4405B
E4405B with
Option UKB
E4407B with
Option UKB
When operating in dc coupled mode, ensure protection of the input
CAUTION
mixer by limiting the input level to 0 Vdc, +30 dBm.
Key Access:
102
Input/Output
set to
Preset Type
Factory
AC
Frequency Range
100 kHz to 3 GHz
100 kHz to 6.7 GHz
100 kHz to 6.7 GHz
100 kHz to 13.2 GHz
100 kHz to 13.2 GHz
10 MHz to 26.5 GHz
Input/Output
). Some amplitude specifications
DC
Frequency Range
100 Hz to 3 GHz
9 kHz to 6.7 GHz
100 Hz to 6.7 GHz
9 kHz to 13.2 GHz
100 Hz to 13.2 GHz
100 Hz to 26.5 GHz
Chapter 2