[:SENSe]:POWer Subsection

Quasi-peak Detector Gain
[:SENSe]:POWer:QPGain[:STATe] OFF|ON|0|1
[:SENSe]:POWer:QPGain[:STATe]?
Turns on or off the linear × 10 gain stage in the quasi-peak and average detector
signal path. This stage may be enabled/disabled whenever the quasi-peak or EMI
average detector is first selected. See SENSe:DETector:FUNCtion:EMI .
Factory Preset
and *RST:
Off
Remarks:
This command is only available with Option AYQ installed.
Front Panel
Access:
Det/Demod
Input Attenuation
[:SENSe]:POWer[:RF]:ATTenuation <rel_ampl>
[:SENSe]:POWer[:RF]:ATTenuation?
Set the input attenuator. This value is set at its auto value if input attenuation is set
to auto.
Factory Preset
and *RST:
10 dB
Range:
ESA E4401B, E4411B: 0 to 60 dB
ESA E4402B, E4403B, E4404B, E4405B: 0 to 75 dB
ESA E4407B, E4408B: 0 to 65 dB
Default Unit:
dB
Front Panel
Access:
AMPLITUDE/Y Scale, Attenuation Auto Man
Input Port Attenuator Auto
[:SENSe]:POWer[:RF]:ATTenuation:AUTO OFF|ON|0|1
[:SENSe]:POWer[:RF]:ATTenuation:AUTO?
Select the input port attenuator range to be set either automatically or manually.