Remote Command Reference
IEEE Common Commands
Front Panel
Access:

Reset

*RST
This command presets the instrument to a factory defined condition that is
appropriate for remote programming operation. *RST is equivalent to performing
the two commands : SYSTem:PRESet and *CLS . This command always
performs a factory preset.
The preset performed by *RST is always a factory preset. That is, the same preset
NOTE
performed by : SYSTem:PRESet when : SYSTem:PRESet:TYPE is set to
FACTory .
Front Panel
Access:
Save
*SAV <register>
This command saves the instrument state to the specified instrument memory
register.
Range:
Remarks:
Front Panel
Access:
238
: MMEMory:STORe:STATe
If the state being loaded has a newer firmware revision than the
revision of the instrument, no state is recalled and an error is
reported.
If the state being loaded has an equal firmware revision than the
revision of the instrument, the state will be loaded.
If the state being loaded has an older firmware revision than the
revision of the instrument, the instrument will only load the
parts of the state that apply to the older revision.
File, Recall State
Preset
Registers are an integer, 0 to 127
See also commands : MMEMory:LOAD:STATe and
: MMEMory:STORe:STATe
File, Save State
Chapter 5