Remote Command Reference
DISPlay Subsystem
:DISPlay:WINDow:TRACe:X[:SCALe]:OFFSet?
Specifies the frequency offset for all frequency readouts such as center frequency,
except that it does not affect marker count.
Factory Preset
and *RST:
0 Hz
Range:
–500 THz to 500 THz
Default Unit:
Hz
History:
Prior to firmware revision A.06.00, the lower range is –3 GHz.
Remarks:
Frequency offset is not available when frequency scale type is
Log ( [:SENSe]:SWEep:SPACing LINear|LOGarithmic ).
Front Panel
Access:
FREQUENCY/Channel, Freq Offset

Display Line Amplitude

:DISPlay:WINDow:TRACe:Y:DLINe <ampl>
:DISPlay:WINDow:TRACe:Y:DLINe?
Defines the level of the display line, in the active amplitude units if no units are
specified.
Factory Preset
and *RST:
2.5 divisions below the reference level
Range:
10 display divisions below the reference level to the reference
level
Default Unit:
Current active units
Front Panel
Access:
Display, Display Line On Off
Display Line On/Off
:DISPlay:WINDow:TRACe:Y:DLINe:STATe OFF|ON|0|1
:DISPlay:WINDow:TRACe:Y:DLINe:STATe?
Turns the display line on or off.
Factory Preset