Front-Panel Key Reference
Display
Allows you to choose fixed or relative limit lines. The fixed (
Limits Fixed Rel
uses the current limit line as a reference with fixed frequency and
amplitude values. The relative (
line value to be relative to the displayed center frequency and reference
level amplitude values. When limit lines are specified with time, rather
than frequency, the
current amplitude values will be relative to the displayed reference
level amplitude, but the time values will always start at the left edge of
the graticule.
As an example, assume you have a frequency limit line. If the limit line
is specified as fixed, entering a limit line segment with a frequency
coordinate of 300 MHz displays the limit line segment at 300 MHz. If
the same limit line table is specified as relative, it is displayed relative
to the analyzer center frequency and reference level. If the center
frequency is at 1.0 GHz, a relative limit line segment with a frequency
coordinate of 300 MHz will display the limit line segment at 1.3 GHz. If
the amplitude component of the relative limit line segment is –10 dB,
then − 10 dB is added to the reference level value to obtain the
amplitude of the given segment (reference level offset included).
A limit line entered as fixed may be changed to relative, and one
entered as relative may be changed to fixed. When changing between
fixed and relative limit lines, the frequency and amplitude values in the
limit line table change so that the limit line remains in the same
position for the current frequency and amplitude settings of the
analyzer. If a time and amplitude limit line is used, the amplitude
values change but the time values remain the same.
Key Access:
Allows you to purge data from the limit-line tables. Pressing
Delete
All Limits
Limits
delete, will delete the limits.
Key Access:
Selects the screen position for the Active Function Display. Depending
Active Fctn
on the type of trace data being viewed, you can move the Active
Position
Function Display position for less visual interference with your screen
data.
Key Access:
Top
Do not select
NOTE
tests as the pass/fail messages for these functions will be overwritten by
the active function information.
78
setting only affects the amplitude values. The
Rel
,
Display
Limits
after the prompt, If you are sure, press key again to
,
Display
Limits
Display
Displays the active function in the top-left corner of the
display.
Active Fctn Position
) setting causes the current limit
Rel
,
when performing limit or margin
Top
) type
Fixed
Delete
Chapter 2