Input Port Power Gain - Agilent Technologies ESA Series User's/Programmer's Reference

Core spectrum analyzer functions
Hide thumbs Also See for ESA Series:
Table of Contents

Advertisement

Remote Command Reference
[:SENSe]:POWer Subsection
On – Input attenuation is automatically set as determined by the Reference
Level Setting.
Off – Input attenuation is manually set
Factory Preset
and *RST:
On
Front Panel
Access:
AMPLITUDE Y Scale, Attenuation

Input Port Power Gain

[:SENSe]:POWer[:RF]:GAIN[:STATe] OFF|ON|0|1
[:SENSe]:POWer[:RF]:GAIN[:STATe]?
Turns the internal preamp on or off.
Factory Preset
and *RST:
Off
Remarks:
This command is only available with Option 1DS installed.
Front Panel
Access:
AMPLITUDE/Y Scale
Input Port Maximum Mixer Power
[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer] <ampl>
[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer]?
Specifies the maximum power at the input mixer.
Factory Preset
− 10 dBm
and *RST:
Range:
–100 dBm to 10 dBm
Default Unit:
dBm
Front Panel
Access:
AMPLITUDE/Y Scale, Max Mixer Lvl
Optimize Preselector Frequency
[:SENSe]:POWer[:RF]:PADJust <freq>
[:SENSe]:POWer[:RF]:PADJust?
This command allows user-defined adjustment of the preselector frequency to
332
,
Int Preamp On Off
Chapter 5

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents