Type Of Averaging For Measurements - Agilent Technologies ESA Series User's/Programmer's Reference

Core spectrum analyzer functions
Hide thumbs Also See for ESA Series:
Table of Contents

Advertisement

Factory Preset
and *RST:
History:
Front Panel
Access:

Type of Averaging for Measurements

[:SENSe]:AVERage:TYPE VIDeo|RMS
[:SENSe]:AVERage:TYPE?
Successive measurements of data can be combined to average out measurement
variations. Detector is set to average and Avg type is set to power (RMS) to
measure RMS voltage (avg power).
NOTE
As a best practice, set amplitude scale ( :DISP:WIND:TRAC:Y:SPAC ) prior to
average type.
VIDeo logarithmically averages the power of the video data (typical units are
dBm). This command is equivalent to pressing front panel keys
Type, Video.
RMS averages the linear power of successive measurements (typical units are
watts).
The following parameters of this command are supported, but not recommended
for new designs. They are provided for limited compatibility to other spectrum
analyzers. When used, the parameters are converted as follows:
TYPE LINear maps to RMS.
TYPE LPOWer maps to VIDeo.
TYPE POWer maps to RMS.
TYPE SCALar and VOLTage will map to VIDeo in linear amplitude scale. If the
amplitude scale is LOG, an error is generated.
TYPE LOG maps to VIDeo. If the amplitude scale is not LOG (linear or Y Axis Units
= Hz), an error is generated.
For compatibility with firmware revisions prior to A.08.00, q
[:SENSe]:AVERage:TYPE? will return LPOW or POW if LPOW or POW is used
during the setting and no further changes have occurred to set the average type (such as
from the front panel).
Factory Preset
and *RST:
Chapter 5
On
Added with firmware revision A.08.00.
BW/Avg, Avg Type, Auto Man
VID
Remote Command Reference
[:SENSe]:AVERage Subsection
BW/Avg, Avg
uery
307

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents