Query The Internal Or External Frequency Reference - Agilent Technologies ESA Series User's/Programmer's Reference

Core spectrum analyzer functions
Hide thumbs Also See for ESA Series:
Table of Contents

Advertisement

Query the Internal or External Frequency Reference

:CALibration:FREQuency:REFerence?
This is a query only. It reports the location of where the instrument frequency
reference is generated.
Range:
Coarse Adjust the Frequency Reference
:CALibration:FREQuency:REFerence:COARse <setting>
:CALibration:FREQuency:REFerence:COARse?
Allows coarse adjustment of the internal 10 MHz reference oscillator timebase of
the analyzer.
NOTE
:CALibration:ALL is required after COARse is set.
Range:
Front Panel
Access:
Fine Adjust the Frequency Reference
:CALibration:FREQuency:REFerence:FINE <setting>
:CALibration:FREQuency:REFerence:FINE?
Allows fine adjustment of the analyzer internal 10 MHz reference oscillator
timebase.
NOTE
:CALibration:ALL is required after FINE is set.
Range:
Front Panel
Access:
Select the Frequency Corrections
:CALibration:FREQuency[:STATe] OFF|ON|0|1
:CALibration:FREQuency[:STATe]?
Turns on or off the frequency corrections.
Chapter 5
INT or EXT
Integer, 0 to 255
System, Alignments, Time Base, Coarse
Integer, 0 to 255
System, Alignments, Time Base, Fine
Remote Command Reference
CALibration Subsystem
267