Agilent Technologies ESA Series User's/Programmer's Reference page 159

Core spectrum analyzer functions
Hide thumbs Also See for ESA Series:
Table of Contents

Advertisement

Freq Correct
On Off
Frequency correction factors must be on for the analyzer to meet its
NOTE
specified performance.
Timebase
Chapter 2
Initiates the alignment of the RF
RF
system. Models E4401B and E4411B
only. This Alignment uses an Internal
Signal Source.
Key Access:
Now
TG
Initiates the alignment of the tracking
(Ext Cable)
generator only when the Option 1DN is
installed in Agilent E4402B, E4403B,
E4404B, E4405B, E4407B, and
E4408B. You must connect a cable from
RF OUT 50 Ω to INPUT 50 Ω
Key Access:
Now
Initiates an alignment of the FM
FM Demod
demodulation system only when Option
BAA or Option 106 is installed.
Key Access:
Now
After a five minute warm-up,
cable connected from AMPTD REF OUT to INPUT 50
Ω ) should be performed on Agilent E4402B, E4403B,
E4404B, E4405B, E4407B, and E4408B. Refer to your
analyzer's appropriate Specifications Guide for details
on using
when
Align Now
Controls use of some of the correction factors. When On
is underlined, frequency correction factors are used.
When Off is underlined, frequency correction factors
are not used, and Freq Corr Off appears on the
display. Turning the correction factors off degrades
frequency accuracy.
Key Access:
,
System
Alignments
Accesses the
and
Fine
which allow control of the 10 MHz time base frequency.
This function is used to verify the settability of the
10 MHz reference and to adjust for changed operating
conditions, such as temperature. Pressing
or cycling power, resets the time base to its original
value.
Front-Panel Key Reference
,
System
Alignments, Align
,
System
Alignments, Align
,
System
Alignments, Align
,
(with a
Align Now
All
is
.
Auto Align
Off
Timebase menu keys,
Coarse
Load Defaults
System (Local)
159