CALCulate:LLINe Subsection
Limit lines can be defined for your measurement. You can then have the instrument
compare the data to your defined limits and indicate a pass/fail condition.
Refer also to :MMEMory and :TRACe subsystems for more trace and limit line
NOTE
commands.
Delete All Limit Lines in Memory
:CALCulate:LLINe:ALL:DELete
Deletes all limit lines in volatile memory.
History:
Front Panel
Access:
Control Limit Line Amplitude Interpolation
:CALCulate:LLINe[1]|2:AMPLitude:INTerpolate:TYPE
LOGarithmic|LINear
:CALCulate:LLINe[1]|2:AMPLitude:INTerpolate:TYPE?
Selects the type of interpolation done for the amplitude values of the designated
limit line when comparing to measured data.
Factory Preset
and *RST:
Remarks:
Front Panel
Access:
Set Fixed or Relative Limit Lines
:CALCulate:LLINe:CMODe FIXed|RELative
:CALCulate:LLINe:CMODe?
Specifies whether the current limit lines are fixed or relative.
NOTE
If you need to change the domain with
: CALCulate:LLINe:CONTrol:DOMain , do it before this command.
Changing the domain deletes all the existing limit line values.
Chapter 5
Added with firmware revision A.08.00.
Display, Limits, Delete All Limits
Not affected by preset
Once this function is defined, the selected type is persistent.
Persistent means that it retains the setting previously selected,
even through a power cycle.
Display, Limits, Limit 1|2, Amptd Interp Log Lin
Remote Command Reference
CALCulate:LLINe Subsection