Remote Command Reference
[:SENSe]:SWEep Subsection

Control Time Gate (Option 1D6 Only)

[:SENSe]:SWEep:TIME:GATE[:STATe] OFF|ON|0|1
[:SENSe]:SWEep:TIME:GATE[:STATe]?
Turns time gating on or off.
Time gate cannot be turned on if external trigger delay is on.
NOTE
Factory Preset
and *RST:
Front Panel
Access:
Time Gate Trigger Type (Option 1D6 Only)
[:SENSe]:SWEep:TIME:GATE:TYPE LEVel|EDGE
[:SENSe]:SWEep:TIME:GATE:TYPE?
Selects between edge and level mode for time-gated spectrum analysis.
Factory Preset
and *RST:
Front Panel
Access:
344
Off
Sweep, Gate, Gate On Off
Level triggers the gate when the signal surpasses a specific level, set to either
low or high.
Edge triggers the gate when the edge of a signal is encountered, set to either a
negative-going edge or a positive-going edge.
Edge
Sweep, Gate, Gate Control Edge Level