For spectrum analyzers with Option 1DN or Option 1DQ, the tracking
NOTE
generator must be turned on (
the segmented sweep table.
When
NOTE
E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B is set to the
maximum attenuation.
Sets the power-sweep function to
Power Sweep
power-sweep range is displayed in the active-function block when you
On Off
press
frequency range when power sweep is on. (To do a power sweep at a
fixed frequency, set the analyzer to zero span at the desired frequency.)
The available power-sweep range is a function of the source attenuator
setting. For your instrument's available power sweep range refer to the
appropriate Specifications Guide for your analyzer.
The output power of the tracking generator is swept according to the
sweep rate of the analyzer. The output power is always swept from the
source power setting to a higher power setting (negative source power
sweep values are not allowed).
Power-sweep measurements are particularly useful in making gain
compression measurements or output power versus frequency
measurements.
When using the power sweep functionality, the actual source amplitude
CAUTION
will be greater than the source amplitude indicated. The source
amplitude value is the starting amplitude for a power sweep, so the
ending amplitude for a power sweep will be higher then the displayed
value. As a result, exercise caution when connecting a power-level
sensitive device to the tracking generator output.
Key Access:
Allows you to select between automatic and manual adjustment of the
Attenuation
tracking generator's output attenuator. The Agilent E4401B and
Auto Man
E4411B can be manually adjusted from 0 to 60 dB in 10 dB steps. All
other Agilent ESA analyzers can be manually adjusted from 0 to 56 dB
in 8 dB steps. When auto-coupled, the attenuation function
automatically adjusts the attenuator to yield the source amplitude level
specified by the
E4401B and E4411B, press
than 10 dB.
Key Access:
Chapter 2
is selected, the output attenuation on the Agilent
Amplitude (Off)
. The analyzer continues to sweep the specified
Power Sweep (On)
Source
Amplitude On Off
Source
,
Source
Amplitude (On)
or
. The value of the
On
Off
softkey function. For the Agilent
for power sweeps greater
Attenuation (Man)
Front-Panel Key Reference
Source
) before you set up
135