Sense]:Detector Subsection - Agilent Technologies ESA Series User's/Programmer's Reference

Core spectrum analyzer functions
Hide thumbs Also See for ESA Series:
Table of Contents

Advertisement

[:SENSe]:DETector Subsection

Automatic Detection Type Selected
[:SENSe]:DETector:AUTO OFF|ON|0|1
[:SENSe]:DETector:AUTO?
Switches automatically to the optimum detection type for typical measurements
using the current instrument settings.
The detector type is average if any of these are on:
• Noise marker
• Band power markers
• Trace averaging when the Average Type is Power (RMS).
The detector type is sample if any of the following conditions are true:
• Trace averaging is on with average type of video
• Both max and min hold trace modes are on
• Resolution bandwidth is less than 1 kHz, and noise marker, band power
markers, or trace averaging is on
• Quasi-peak detection (Option AYQ)
• EMI Average detection (Option AYQ)
The detector type is negative peak if any trace is in min hold and no traces are in
max hold.
The detector type is peak if the above conditions are off.
Manually changing the detector function turns Auto off.
Refer to
Figure 2-1 on page
is determined.
Factory Preset
and *RST:
On
History:
Added with firmware revision A.08.00. Modified with A.14.00.
Front Panel
Access:
Det/Demod
Type of Detection
[:SENSe]:DETector[:FUNCtion]
NEGative|POSitive|SAMPle|AVERage|RMS
[:SENSe]:DETector[:FUNCtion]?
Specifies the detection mode. For each trace interval (bucket), average detection
Chapter 5
57, which shows a decision tree of how detection type
,
Detector
Remote Command Reference
[:SENSe]:DETector Subsection
319