Select Application - Agilent Technologies ESA Series User's/Programmer's Reference

Core spectrum analyzer functions
Hide thumbs Also See for ESA Series:
Table of Contents

Advertisement

Example:
Factory Preset:
Range:
Front Panel
Access:

Select Application

:INSTrument[:SELect] 'SA'|'GSM'|'CDMA'|'PNOISE'|'BLUETOOTH'
|'EDGE'|'MAN'|'LINK'|'CATV'|'NFIGURE'
:INSTrument[:SELect]?
Select the measurement mode. The actual available choices depend upon which
modes (measurement applications) are installed in the instrument. A list of the
valid choices is returned with the INST:CAT? query.
Once an instrument mode is selected, only the commands that are valid for that
mode can be executed.
If you are using the SCPI status registers and the analyzer mode is changed, the
NOTE
status bits should be read, and any errors resolved, prior to switching modes. Error
conditions that exist prior to switching modes cannot be detected using the
condition registers after the mode change. This is true unless they recur after the
mode change, although transitions of these conditions can be detected using the
event registers.
Changing modes resets all SCPI status registers and mask registers to their
power-on defaults. Hence, any event or condition register masks must be
re-established after a mode change. Also note that the power up status bit is set by
any mode change, since that is the default state after power up.
SA (Spectrum Analysis)
GSM
CDMA (cdmaOne)
PNOISE (phase noise)
EDGE
NFIGURE (noise figure)
BLUETOOTH
CATV (Cable TV)
MAN (Modulation Analysis)
LINK (89600 VSA Link Software)
Chapter 5
INST:NSEL 4
Persistent state with factory default of
1 to x, where x depends upon which applications are installed.
MODE
Remote Command Reference
INSTrument Subsystem
295