Turn Automatic Averaging On/Off

[:SENSe]:AVERage:TYPE:AUTO OFF|ON|0|1
[:SENSe]:AVERage:TYPE:AUTO?
Sets the averaging to be automatically set to the appropriate type for the current
measurement setup. Or allows you to manually choose the type of averaging with
[:SENSe]:AVERage:TYPE .
When AUTO is On:
See
Figure 5-1
Auto Rules for Average Type
306
If the Y Axis Scale is not Linear or Log, then average type is Video (Y Axis
Scale) Averaging.
If the Y Axis Scale is Linear or Log, then average type is Power Averaging.
If the Detector is Peak, Sample, or Negative Peak (not Average), then average
type is Video Average.
Figure
5-1, which shows these auto rules for average type in flowchart format.