Remote Command Reference

[:SENSe]:SIDentify Subsection

[:SENSe]:SIDentify Subsection
Set Mixer Signal Identification Mode
[:SENSe]:SIDentify:MODE ISUPpress|ISHift
[:SENSe]:SIDentify:MODE?
This command lets you choose one of two types of signal identification methods
when viewing multiple responses from non-preselected external mixers. This
command is available only with Option AYZ (External Mixing). Refer to
:INPut:MIXer:TYPE to select the type of mixer being used.
Factory Preset
and *RST:
ISUPpress
History:
Added with firmware revision A.03.00.
Front Panel
Access:
Input/Output (or Input), Input Mixer, Signal ID Mode, Image
Suppress
Input/Output (or Input), Input Mixer, Signal ID Mode, Image
Shift
Set Mixer Signal Identification State
[:SENSe]:SIDentify[:STATe] OFF|ON|0|1
[:SENSe]:SIDentify[:STATe]?
This command activates an algorithm which either removes or aids with the
identification of multiple responses. These responses are generated from a single
input signal using non-preselected external mixers.
Factory Preset
and *RST:
Off
History:
Added with firmware revision A.03.00.
Remarks:
Signal identification relies on the acquisition of data from two
successive sweeps. Therefore, if the analyzer is in single sweep
mode, two sweep triggers are needed to generate the sweep pair.
In image suppress mode, synchronization is ensured by first
turning off signal identification, initiating a single sweep, then
turning on signal identification followed by two single sweeps.