Remote Command Reference
CALibration Subsystem

Automatic Alignment

:CALibration:AUTO OFF|ON|0|1
:CALibration:AUTO?
Turns the automatic alignment on and off. This is run continuously, at the
completion of each sweep.
Factory Preset
and *RST:
On at power-up
Front Panel
Access:
System, Alignments, Auto Align, All
System, Alignments, Auto Align, All but RF
System, Alignments, Auto Align, Off
Return to the Default Alignment Data
:CALibration:DATA:DEFault
Initializes the alignment data to the factory defaults.
Front Panel
Access:
System, Alignments, Load Defaults
Align FM Demodulation
:CALibration:FMDemod
:CALibration:FMDemod?
Performs an alignment of the FM demodulation hardware if Option 106, AYQ or
BAA is installed. The query form of this command performs the alignment and
returns zero if the alignment is successful. If you have multiple FM demodulation
hardware options, the hardware that will be used is automatically selected in order
of priority (with firmware revisions A.14.00 or later).
1. Option 106 - Bluetooth 
2. Option AYQ - Quasi-peak detection and FM demodulation
3. Option BAA - FM Demodulation
Front Panel
Access:
System, Alignments, Align Now, FM Demod
1. Bluetooth is a trademark owned by its proprietor and used by Agilent Technologies under