Set TV Field Mode

:TRIGger[:SEQuence]:TV:FMODe ENTire|ODD|EVEN
:TRIGger[:SEQuence]:TV:FMODe?
This command allows the user to determine how the fields of the TV picture signal
will be treated by the trigger system. Requires Option BAA and B7B.
Factory Preset
and *RST:
Range:
NOTE
Refer to
command dependencies.
History:
Front Panel
Access:
Chapter 5
ENTire (entire frame)
ENTire (entire frame)
For formats NTSC-M, NTSC-Japan, and PAL-M, the minimum
line is 1, and the maximum line is 525.
For formats PAL-B, D, G, H, I, PAL-N, PAL-N Combin, and
SECAM-L, the minimum line is 1, and the maximum line is
625.
ODD (Field 1)
For formats NTSC-M, NTSC-Japan, and PAL-M, the minimum
line is 1, and the maximum line is 263.
For formats PAL-B, D, G, H, I, PAL-N, PAL-N Combin, and
SECAM-L, the minimum line is 1, and the maximum line is
313.
EVEN (Field 2)
For formats NTSC-M, NTSC-Japan, and PAL-M, the minimum
line is 1, and the maximum line is 262.
For formats PAL-B, D, G, H, I, PAL-N, PAL-N Combin, and
SECAM-L, the minimum line is 1, and the maximum line is
312.
"TV" on page 171
for a more detailed explanation of TV fields and
Added TV trigger mode with firmware revision A.03.00.
Trig, TV Trig Setup, Field, Entire Frame
Trig, TV Trig Setup, Field, Field One
Trig, TV Trig Setup, Field, Field Two
Remote Command Reference