[:SENSe]:SWEep Subsection

[:SENSe]:SWEep Subsection
Sweep Points
[:SENSe]:SWEep:POINts <number of points>
[:SENSe]:SWEep:POINts?
This command sets the number of sweep points.
Factory Preset
and *RST:
401
Example:
:SWEep:POIN 401
History:
This command is available only on analyzers with firmware
revision A.04.00 and later. Analyzers with firmware revisions
prior to A.04.00 have the number of sweep points fixed at 401.
Range:
101 to 8192, (2 to 8192 in zero span for analyzers with firmware
revision A.05.00 and later)
Remarks:
The number of sweep points can be set only with Agilent ESA
models E4401B, E4402B, E4404B, E4405B, and E4407B.
Sweep points for ESA models E4403B, E4408B, and E4411B
are fixed at 401 points. The number of sweep points may be
queried on all analyzers with firmware revision A.04.00 and
later.
For analyzers with firmware revisions prior to A.08.00, any
change to sweep points sets the following commands as shown:
:CALCulate:LLINe1:DISPlay to off, and
:CALCulate:LLINe2:DISPlay to off.
Whenever the number of sweep points change, the following
functions are affected:
• All trace data is erased
• Any traces in view mode will go to blank mode
• Sweep time is re-calculated
• Any limit lines that are on will be turned off (For analyzers
with firmware revisions prior to A.08.00)
Front Panel
Access:
Sweep, Points