Front-Panel Key Reference
AMPLITUDE Y Scale
The amplitude correction entered for the lowest frequency will be
NOTE
applied to all frequencies less than the lowest frequency entered.
Similarly, the amplitude correction for the highest frequency entered
will be applied to all frequencies greater than the highest frequency
entered.
For amplitude-correction factors, a maximum of two entries with the
NOTE
same frequency are valid. Only the first and last points of a series with
the same frequency values are used; any middle points are ignored.
Amplitude-correction data is sorted in the table by frequency. The
NOTE
sorting occurs immediately after you have entered the frequency value
via the front-panel.
46
Key Access:
AMPLITUDE Y Scale
,
Corrections
Cable
Point
Allows you to create or edit an
amplitude-correction factor data point.
Up to 200 points may be defined for
each set. Enter the point number to be
created or edited by using the numeric
keypad, then press
Enter
knob, tab, or step keys to move to an
existing point. Press
Bk Sp
errors. After selecting a point,
becomes active.
Frequency
Key Access:
AMPLITUDE Y Scale
,
,
Corrections
Cable
Edit
Frequency
Allows you to enter the frequency value
for an amplitude-correction point.
Enter the frequency value by using the
numeric keypad. Change the frequency
value by using the step keys or the
knob. Press
to correct errors.
Bk Sp
After selecting a point,
becomes active.
A frequency coordinate must always be
specified for amplitude-correction
factors.
Key Access:
AMPLITUDE Y Scale
,
,
Corrections
Cable
Edit
,
More,
, or use the
to correct
,
More,
Amplitude
,
More,
Chapter 2