Remote Command Reference

[SENSe]:MIXer Subsection

[SENSe]:MIXer Subsection
Select External Mixer Band
[:SENSe]:MIXer:BAND K|A|Q|U|V|E|W|F|D|G|Y|J|USER
[:SENSe]:MIXer:BAND?
This command allows the selection of one of the pre-defined bands corresponding
to the external mixer currently in use. This command is available only with Option
AYZ (External Mixing).
Factory Preset
and *RST:
Remarks:
NOTE
Bands K, E, W, F, D, G, Y, and J are not available if Mixer Type is set to Presel.
Front Panel
Access:
External Mixer Bias Adjust
[:SENSe]:MIXer:BIAS <numeric>
[:SENSe]:MIXer:BIAS?
This command allows the adjustment of an internal bias source for use with
external mixers. This command is available only with Option AYZ (External
Mixing).
Factory Preset
and *RST:
Range:
Default Unit:
Remarks:
NOTE
Mixer Bias will be set to Off when transitioning from manual harmonic mode to
auto harmonic mode.
Front Panel
Access:
328
Band A (26.5-40 Ghz)
If the mixer harmonic configuration
( [:SENSe]:MIXer:HARMonic:AUTO OFF|ON|0|1 is
set to Off (manual), then a query will return " USER ".
Input/Output (or Input), Input Mixer, Ext Mix Band
0
–10 mA to 10 mA
mA
The bias signal is present on the center conductor of the IF
INPUT connector on the front panel. See related command
[:SENSe]:MIXer:BIAS[:STATe] OFF|ON|0|1 .
Input/Output (or Input), Input Mixer, Mixer Config, Mixer
Bias On Off