Video Trigger Level Amplitude - Agilent Technologies ESA Series User's/Programmer's Reference

Core spectrum analyzer functions
Hide thumbs Also See for ESA Series:
Table of Contents

Advertisement

Remote Command Reference
TRIGger Subsystem
Example:
History:
Remarks:
Front Panel
Access:

Video Trigger Level Amplitude

:TRIGger[:SEQuence]:VIDeo:LEVel <ampl>
:TRIGger[:SEQuence]:VIDeo:LEVel?
Specifies the level at which a video trigger will occur.
Factory Preset
and *RST:
Range:
Default Unit:
Remarks:
NOTE
Trigger Delay is not available in Video trigger mode, so turning Video on turns off
Trigger Delay, but preserves the value of Trigger Delay.
Front Panel
Access:
388
:TRIG:SEQ:TV:STAN MNTS
Added TV trigger mode with firmware revision A.03.00.
Once this function is defined, the selected type is persistent.
Persistent means that it retains the setting previously selected,
even through a power cycle. The setting will change with "Load
State."
Trig, TV Trig Setup, Standard, NTSC-M
Trig, TV Trig Setup, Standard, NTSC-Japan
Trig, TV Trig Setup, Standard, PAL-M
Trig, TV Trig Setup, Standard, PAL-B,D,G,H,I
Trig, TV Trig Setup, Standard, PAL-N
Trig, TV Trig Setup, Standard, PAL-N Combin
Trig, TV Trig Setup, Standard, SECAM-L
2.5 divisions below reference level
10 display divisions below reference level to reference level
current amplitude units
Video is adjusted using this command, but must also be selected
using the command :TRIGger[:SEQuence]:SOURce
VIDeo . When in FM Demod and Demod View is on, the Video
Trigger level is adjusted/queried using the command
:TRIGger[:SEQuence]:VIDeo:LEVel:FREQuency
<freq> .
Trig, Video