Sweep - Agilent Technologies ESA Series User's/Programmer's Reference

Core spectrum analyzer functions
Hide thumbs Also See for ESA Series:
Table of Contents

Advertisement

Sweep

Activates the sweep time function and accesses the following menu
keys:
Selects the length of time the analyzer takes to tune across the
Sweep Time
displayed frequency span (or, in zero span, the time the analyzer takes
Auto Man
to sweep the full screen). Reducing the sweep time increases the sweep
rate. The sweep time can be changed using the step keys, the knob, or
the numeric keypad.
In non-zero spans:
When the sweep time is auto-coupled, the analyzer selects the optimum
(shortest) sweep time for the current settings. This selection process is
influenced by several factors:
• The maximum tuning rate of the spectrum analyzer
• The selected Resolution Bandwidth and Video Bandwidth filters
• The maximum sample rate of the Analog-to-Digital converter (ADC)
(Trace data are digitized and stored in memory)
• The number of trace points (
• The Amplitude Scale (logarithmic or linear)
• The Detector mode
You may select a sweep time less than the auto-coupled value; however
this may generate measurement errors. If this happens, the error
message: Meas Uncal will appear in the upper right corner of the
display. When the
is linear, the fastest sweet time is 8 milliseconds.
In zero span:
The minimum sweep time is determined by the maximum sample rate
of the ADC and the number of sweep points. With Option B7D, the
fastest sweep time is 2.5 microseconds (with 101 sweep points). As the
number of points increases, the sweep time increases. With Option
AYX, the fastest sweep time is 5 microseconds (with 101 sweep points).
With neither option installed, the fastest sweep time is 1 millisecond
(with 101 sweep points) and 4 milliseconds (with 401 sweep points).
When the
the fastest sweet time is 8 milliseconds.
With software A.08.00 and greater, in zero span, the auto/manual
NOTE
function of this key is not applicable. When
selected in non-zero span, any changes to Sweep Time while in zero
span will revert to the Auto value when you return to non-zero span.
When
Sweep Time while in zero span will be maintained when you return to
non-zero span.
Chapter 2
detector is selected and the amplitude scale
Average
detector is selected and the amplitude scale is linear,
Average
is selected in non-zero span, any changes to
Sweep Time (Man)
Front-Panel Key Reference
)
Sweep, Points
Sweep Time (Auto)
Sweep
is
145