Source - Agilent Technologies ESA Series User's/Programmer's Reference

Core spectrum analyzer functions
Hide thumbs Also See for ESA Series:
Table of Contents

Advertisement

Front-Panel Key Reference

Source

Source
Accesses the tracking generator key functions (Options 1DN or 1DQ
only). Without Option 1DN or 1DQ, pressing
message Option not installed to appear on the analyzer display.
If
CAUTION
Auto Align
retuned to approximately 1.557 GHz between most sweeps. Some
devices under test (for example, amplifiers with AGC) may be
susceptible to damage due to this momentary retuning. To avoid this
momentary retuning, turn the auto align off by pressing
Alignments
instrument to learn more information about using the analyzer with
Auto Align set to off.
The output frequency of the 3 GHz tracking generator may not exactly
NOTE
match the input frequency of the spectrum analyzer. This is probably
not a problem when using wide resolution bandwidths. However,
narrowing the RBW may cause a drop in the displayed signal amplitude
as the frequency offset puts the signal out of the center of the RBW
filter. For narrower RBWs you may need to do a tracking peak to adjust
the tracking generator frequency. See the
Tracking Adjust
Activates (
Amplitude
generator. The power level can then be adjusted using the numeric
On Off
keypad, step keys, or knob. Pressing any digit, 0 through 9, on the
numeric keypad brings up the selected terminator menu. See the
specifications chapter in the calibration guide for the available output
power for your tracking generator.
Key Access:
Power-level sensitive devices connected to the tracking generator
CAUTION
output may be accidentally damaged. This is because the actual source
amplitude will be greater than the amplitude indicated on the analyzer,
when the power sweep function is being used.
The source amplitude value is used as the starting amplitude for a
power sweep. The ending amplitude for a power sweep is higher then
the source amplitude value and it may actually be set to values beyond
specified output levels. Therefore, exercise caution when connecting a
power-level sensitive device to the tracking generator output.
134
is on, the 3 GHz tracking generator will be momentarily
,
,
. Refer to the Specifications Guide for your
Auto Align
Off
.
) or deactivates (
On
Off
Source
will cause the error
Source
Tracking Peak
) the output power of the tracking
,
System
and
Man
Chapter 2

Hide quick links:

Advertisement

Table of Contents
loading