Agilent Technologies ESA Series User's/Programmer's Reference page 217

Core spectrum analyzer functions
Hide thumbs Also See for ESA Series:
Table of Contents

Advertisement

For example, to enable bit 9 and bit 3 so that whenever either of those bits is set to 1, the Questionable
Status Summary bit of the Status Byte Register will be set to 1, send the command
:STAT:QUES:ENAB 520 (512 + 8). The command :STATus:QUEStionable:ENABle? returns
the decimal value of the sum of the bits previously enabled with the
:STATus:QUEStionable:ENABle <num> command.
Figure 4-11
Questionable Status Event Enable Register
Bit descriptions in the Status Questionable Condition Register are given in the following table.
Bit
Chapter 4
Use Status Registers to Determine the State of Analyzer Events and Conditions
Decimal
Description
Value
0
1
Reserved: This bit is not used by the analyzer, but are for
future use with other Agilent products.
1
2
Reserved: This bit is not used by the analyzer, but are for
future use with other Agilent products.
2
4
Reserved: This bit is not used by the analyzer, but are for
future use with other Agilent products.
3
8
POWer Summary: This is the summary bit for the
Questionable Power Status Register.
4
16
Reserved: This bit is not used by the analyzer, but are for
future use with other Agilent products.
5
32
FREQuency Summary: This is the summary bit for the
Questionable Frequency Status Register.
6
64
Reserved: This bit is not used by the analyzer, but are for
future use with other Agilent products.
7
128
Reserved: This bit is not used by the analyzer, but are for
future use with other Agilent products.
Status Registers
217

Hide quick links:

Advertisement

Table of Contents
loading