Sets The Source Output Power Mode - Agilent Technologies ESA Series User's/Programmer's Reference

Core spectrum analyzer functions
Hide thumbs Also See for ESA Series:
Table of Contents

Advertisement

Factory Preset
− 10 dBm
and *RST:
Range:
ESA E4401B, E4411B with option 1DN (50 Ohm Tracking
Generator): –70 dBm to 3 dBm
ESA E4401B, E4411B with Option 1DQ (75 Ohm Tracking
Generator): –27.25 dBmV to 45.75 dBmV
ESA E4402B, E4403B: –66 dBm to 3 dBm
ESA E4404B: –66 dBm to 3 dBm
ESA E4405B: –66 dBm to 3 dBm
ESA E4407B, E4408B: –66 dBm to 3 dBm
Default Unit:
dBm
Front Panel
Access:
Source, Amplitude On Off

Sets the Source Output Power Mode

:SOURce:POWer:MODE FIXed|SWEep
:SOURce:POWer:MODE?
Sets the source output to be at a single amplitude (fixed) or to sweep through a
range of power levels.
Factory Preset
and *RST:
Fixed
Front Panel
Access:
Source, Power Sweep On Off
Set the Source Sweep Power Range
:SOURce:POWer:SPAN <rel_ampl>
:SOURce:POWer:SPAN?
Specifies the range of power levels through which the source output will sweep.
Use : SOURce:POWer:STARt to set the power level at the start of the power
sweep. This command is equivalent to : SOURce:POWer:SWEep.
Factory Preset
and *RST:
0 dB
Range:
0 dB to 20 dB
Default Unit:
dB
Chapter 5
Remote Command Reference
SOURce Subsystem
347

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents