Remote Command Reference
CALibration Subsystem
Calibrate the Tracking Generator
:CALibration:TG
:CALibration:TG?
Performs an alignment of the tracking generator assembly.
The query performs the alignment and returns a zero if the alignment is successful.
NOTE
This command is only applicable on Agilent ESA models E4402B, E4403B,
E4404B, E4405B, E4407B, and E4408B. Before executing this command, connect
a cable between front panel connector
and the
connector. The
RF OUT
INPUT
alignment will fail using command CAL:TG if the cable is not connected.
Front Panel
Access:
System, Alignments, Align Now, TG
Chapter 5
269