Remote Command Reference
CALibration Subsystem
Factory Preset
and *RST:
Front Panel
Access:

Align the RF Circuitry

:CALibration:RF
:CALibration:RF?
Performs an alignment of the RF assembly.
The query performs the alignment and returns a zero if the alignment is successful.
Before executing this command, connect a cable between front panel connector
AMPTD REF OUT
analyzers except Agilent models E4401B and E4411B. If the cable is not
connected, the alignment will fail.
Front Panel
Access:
Select the Source State for Calibration
:CALibration:SOURce:STATe OFF|ON|0|1
:CALibration:SOURce:STATe?
Controls the state of the 50 MHz alignment signal.
The alignment signal is internally switched to the
NOTE
E4401B and E4411B. For all other models, connect a cable between front panel
connector
calibration.
Factory Preset
and *RST:
Front Panel
Access:
268
On
System, Alignments, Freq Correct On Off
and the
INPUT
connector for all Agilent ESA spectrum
System, Alignments, Align Now, RF
AMPTD REF OUT
and the
Off
For Agilent ESA models E4401B and E4411B:
Input/Output (or Input), Amptd Ref (f=50 MHz) On Off
For all other Agilent ESA models:
Input/Output (or Input), Amptd Ref Out (f=50 MHz) On Off
for Agilent models
INPUT
INPUT
connector before performing a