Initiate:continuous; Initiate[:Immediate] - Keysight Technologies E4981A Programming Manual

120 hz/1 khz/1 mhz capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Command Reference
E4981A SCPI Commands

:INITiate:CONTinuous

Syntax
:INITiate:CONTinuous {ON|OFF|1|0}
:INITiate:CONTinuous?
Description
This command changes the trigger state from "idle" state to the "wait for trigger"
automatically or stay in "idle" state. After executing :SYSTem:PRESet, this value changes
to ON, after
For details on the trigger system, refer to
NOTE
This command is turned ON when the trigger source is changed from the front panel of the
E4981A.
Parameters
ON or 1
OFF or 0 (initial value)
This setup is initialized to ON when using the
when using the
Query response
{1|0}<newline><^END>
Related commands
*RST on page 167
:SYSTem:PRESet on page 279
Equivalent key
No equivalent key is available on the front panel.
sequence

:INITiate[:IMMediate]

Syntax
:INITiate[:IMMediate]
Description
Changes the trigger state from "idle" state to "wait for trigger" state for one trigger
sequence. (No Query)
For details on the trigger system, refer to
Related commands
:INITiate:CONTinuous on page 224
Equivalent key
No equivalent key is available on the front panel.
sequence
224
*RST
this value changes to OFF .
Enables successive startup.
Disables successive startup.
*RST
command.
"Trigger system" on page
Description
:SYSTem:PRESet
command and to OFF
"Trigger system" on page
60.
60. (No query)
Chapter 10

Advertisement

Table of Contents
loading

Table of Contents