Sense Commands - Keysight Technologies BT2152A Operating And Service Manual

Keysight bt2152a self-discharge analyzer
Hide thumbs Also See for BT2152A:
Table of Contents

Advertisement

Sense Commands

SENSe:OCV:AVAIlable?
Returns the availability of the open-circuit voltage measurement. This command is only applicable
when used in conjunction with the
A one (1) is returned when the OCV measurement is completed. A zero (0) is returned when the OCV
measurement is unavailable or if the command was used with any other INIT command besides
INITiate:TEST:OCV.
Parameter
(none)
Returns the availability of OCV : SENS:OCV:AVAI?
SENSe:PROBecheck:AVAIlable?
Returns the availability of the probe check test.
A one (1) is returned when the probe check command is completed. A zero (0) is returned when the
probe check command is unavailable.
Parameter
(none)
Returns the availability of the Probe check test: SENS:PROB:AVAI?
SENSe:THERmistor:CONFigure <temp>, <resistance>, <beta>
SENSe:THERmistor:CONFigure?
Specifies the parameters for the external thermistor connected to the rear panel temperature sensor
input (see
Thermistor
Connections).
<temp> the nominal temperature of the thermistor in degrees C, from -272.15° C to 273.15° C
<resistance> the nominal resistance in ohms, from 1 Ω to 1,000,000 Ω
<beta> value from 1 to 1,000,000
Specifies the nominal temperature and resistance: SENS:THER:CONF 25, 10000, 4073
Keysight BT2152A Operating and Service Guide
INITiate:TEST:OCV
command.
Parameter
4 Programming Reference
Typical Return
0 or 1
Typical Return
0 or 1
Typical Return
25° C
10,000
4073
73

Advertisement

Table of Contents
loading

Table of Contents