3
Calibrations and Test Procedures

Group Parameters

Parameter Name
0 V Amplitude at Init
Reduced Amplitude at Init
Re-Init Sequence
74
Group parameters are used for several related calibration or test
procedures. They are shown on the right side of the N5990A user interface
when the selected entry of the procedure tree on the left is a group instead
of an individual procedure.
The MIPI M-PHY Receiver Test Software has some group parameters (in
addition to Repetitions on the top-level entry of the procedure tree (see
Figure
37). These will be common for all N5990A procedures.
Figure 37
MPHY Group Parameters
Table 5
describes the group parameters.
Table 5
Group Parameters
Parameter Description
The value is True/False. If is set to true, the amplitude of the generator will be set to 0 Volts during the
Init phase of each test.
The voltage value set during the Init phase of each test.
The value is True/False. The default value is false. If the value is set to True, the test sequence starts
again from the beginning in each test step to ensure that the DUT is in a defined state. This results in an
increased test duration
Keysight N5990A MIPI M-PHY Receiver Test User Guide