Procedure
Connection Diagram
Parameters
Dependencies
Keysight N5990A MIPI M-PHY Receiver Test User Guide
This test is similar to the previous Test 2.1.8 but in this case, the data
generator sends Continuous-Mode CJTPAT signaling instead of
Burst-Mode, and the DUT is configured to receive the Continuous-Mode
CJTPAT in Terminated mode. For more details refer to the Procedure
section of the Test 2.1.8-HS-RX Frequency Offset Tolerance (fOFFSET-RX)
during HS Burst).
Refer to the connection diagrams shown in the Connection Diagram
section of the Test 2.1.1- HS-RX Differential Input Voltage Amplitude Test.
The parameters used in Expert Mode for this test, are as follows:
•
CDR Loop Bandwidth (if CDR is selected in the Special Parameter
dialog, see
"BERT Analyzer
•
CDR Peaking (if CDR is selected in the Special Parameter dialog, see
"BERT Analyzer
Settings")
•
CDR Transition Density (if CDR is selected in the Special Parameter
dialog, see
"BERT Analyzer
•
CDR Loop Order (if CDR is selected in the Special Parameter dialog,
see
"BERT Analyzer
•
Target BER
•
BerReader Init Mode
•
Test Sequence
•
Lane Under Test Termination Model
•
Level Pairs
•
Tested Offset
•
Tested Range
These parameters are listed in alphabetical order in
The calibrations required for this test are as follows:
•
Reference Clock Calibration
•
Level Calibration Terminated
Settings")
Settings")
Settings")
HS Tests
5
Table
6.
193